Key Features & Specifications
Fluke - series II - 5720A - High Performance Multifunction Calibrator
- Minimum output: 0 V for all ranges, except 100 V for 1100 V range
- Maximum load: 50 mA for 2.2 V through 220 V ranges; 20 mA for 1100 V range; 50 Ω output impedance on 220 mV range; all ranges < 1000 pF, > 25 Ω
- Load regulation: < (0.2 ppm of output + 0.1 ppm of range), full load to no load
- Line regulation: < 0.1 ppm change, ± 10 % of selected nominal line
- Settling time: 3 seconds to full uncertainty; + 1 second for range or polarity change; + 1 second for 1100 V range
- Overshoot: < 5 % Common mode rejection: 140 dB, DC to 400 Hz
Remote sensing: Available 0 V to ± 1100 V, on 2.2 V through 1100 V ranges
Specifications
See datasheet for specifications
Description
The Fluke 5720A set even higher standards, calibrating your high-performance workload up to 8 ½ digit system DMMs. This unit now states specifications to the standard Fluke 99% confidence level as well as the 95% confidence level to support easy measurement comparisons according to international quality standards including ISO 9000 and ISO Guide 25. All specifications are absolute and include the uncertainty of the calibration standards used. No additional analysis is required.